Description
1,Product Introduction
Professional portable conductivity meter, selectable TDS, salinity and resistivity measurement modes, 1 to 5 points calibration, setup menu contains 12 options, accuracy: 0.5% F.S.
	
2,Features
	• Professional portable conductivity meter is equipped with a backlit LCD display.
• 1 to 5 points calibration with automatic recognition for conductivity standards.
• Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear and pure water compensations, seawater and practical salinity measurement modes.
• Automatic temperature compensation corrects conductivity measurements to the reference temperature.
• Auto-read function senses and locks the measurement endpoint.
• Calibration due alarm reminds the user to calibrate the meter regularly.
• Calibration log shows the calibration points and factor.
• Setup menu allows setting the number of calibration points, stability criteria, temperature unit,date and time, etc.
• Reset function automatically resumes all settings back to the factory defaults.
• Expanded memory stores or recalls up to 500 data sets.
• USB communication interface is easy to transfer data to a PC.
• Multi-mode power scheme (batteries, power adapter, USB port) ensures the smooth operation of the meter.
	
3,Specification
	
| Model |  | 540-S | 
| Conductivity |  |  | 
| Range |  | 0 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm | 
| Resolution |  | 0.001, 0.01, 0.1, 1, automatic | 
| Accuracy |  | ±0.5% F.S. | 
| Calibration |  | 1 to 5 points | 
| Calibration Solutions |  | 10 µS / 84 µS / 1413 µS / 12.88 mS / 111.8 mS | 
| TDS |  |  | 
| Range | 
 | 0 to 10.00, 100.0, 1000 ppm, 10.00, 200.0 ppt | 
| Resolution | 
 | 0.01, 0.1, 1, automatic | 
| Accuracy | 
 | ±1% F.S. | 
| TDS Conversion Factor | 
 | 0.1 to 1.0 (default 0.5) | 
| Salinity |  |  | 
| Range |  | 0.00 to 80.00 ppt / 0.00 to 42.00 psu | 
| Resolution |  | 0.01 | 
| Accuracy |  | ±1% F.S. | 
| Resistivity |  |  | 
| Range |  | 0.00 to 20.00 MΩ | 
| Resolution |  | 0.01 | 
| Accuracy |  | ±1% F.S. | 
| Temperature |  |  | 
| Range |  | 0 to 105°C / 32 to 221°F | 
| Resolution |  | 0.1°C/°F | 
| Accuracy |  | ±0.5°C / ±0.9°F | 
| Calibration |  | 1 point | 
| Calibration Range |  | Reading ±10°C/°F | 
| Other Specifications |  |  | 
| Temperature Compensation |  | 0 to 100°C / 32 to 212°F, manual or automatic | 
| Temperature Coefficient | 
 | 0.0 to 10.0%/°C | 
| Pure Water Compensation | 
 | Yes | 
| Reference Temperature | 
 | 20 / 25°C | 
| Cell Constant | 
 | K=0.1 / 1 / 10 / Custom | 
| Hold Function | 
 | Manual / Auto-read measurement endpoint | 
| Stability Criteria | 
 | Low / High | 
| Calibration Due Alarm | 
 | 1 to 31 days or off | 
| Auto-Power Off | 
 | 10 / 20 / 30 minutes after last key pressed | 
| Memory | 
 | Stores up to 500 data sets | 
| Output | 
 | USB communication interface | 
| Connector | 
 | 6-pin mini-DIN, 3.5 mm jack socket | 
| Display | 
 | Custom LCD (80 × 60 mm) | 
| Power Supply | 
 | 3 × 1.5V AA batteries or 5V DC power adapter | 
| Battery Life |  | Approximately 150 hours without backlight | 
| Environmental Conditions |  | 0 to 50°C / 32 to 122°F, relative humidity < 80% | 
| Dimensions |  | 170 × 85 × 30 mm | 
| Weight |  | 300 g | 
 
 
 
 
 
 
 